Appendix C In Situ, Derated,
and Water Value Intensities
All intensity parameters are measured in water. Since water absorbs very little acoustic
energy, these water measurements represent a worst case value. Biological tissue does
absorb acoustic energy. The true value of the intensity at any point depends on the amount
and type of tissue and the frequency of the ultra-sound that passes through the tissue. The
intensity value in the tissue, In Situ, has been estimated by using the following formula:
In Situ = Water[e
-0:23alf
]
where:
In Situ = In Situ Intensity Value
Water = Water Value Intensity
e = 2.7183
a = Attenuation Factor
Tissue = a (dB/cm-MHz)
Amniotic Fluid = 0.006
Brain = 0.53
Heart = 0.66
Kidney = 0.79
Liver = 0.43
Muscle = 0.55
l = Skin line to measurement depth (cm)
f = Center frequency of the transducer/system/mode combination (MHz)
Since the ultrasonic path during an examination is likely to pass through varying lengths
and types of tissue, it is difficult to estimate the true In Situ intensity. An attenuation
factor of 0.3 is used for general reporting purposes; therefore, the In Situ value which is
commonly reported uses the formula:
In Situ derated = Water[e
-0.069lf
]
Since this value is not the true In Situ intensity, the term “derated” is used.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...