![MediSono P25 EXP User Manual Download Page 40](http://html1.mh-extra.com/html/medisono/p25-exp/p25-exp_user-manual_1762413040.webp)
3 Preparing the System
28
Basic User Manual
3
�
3
�
2 Using the Battery
!
●
Do not disassemble or alter the battery. Otherwise, there is a danger of explosion.
●
Replace the battery with the same or the equivalent type.
●
Do not short-circuit the battery by directly connecting the system with metal objects.
●
If the battery emits an odor or heat, is deformed or discolored, or appears abnormal
during use, charging or storage, remove it and stop using it immediately. If you have
any questions about the battery, please contact the manufacturer.
●
Do not use a battery if it leaks. If your skin or clothing is stained with the fluid from
the battery, thoroughly wash the area immediately with clean water. If the fluid comes
into contact with your eyes, immediately flush your eyes with water and seek an
oculist for help.
●
If the battery is found damaged, severely degraded or used for 3 years, dispose of it in
accordance with the local laws or regulations.
If the system is connected to the mains supply through a power supply cable, the battery is
automatically charged.
If the system is disconnected from the mains supply, the system is automatically powered
by the battery.
●
A fully charged battery can work continuously for more than one hour.
●
Two batteries are used for system power supply, they must satisfy the following
specifications,
Nominal voltage: 14.8V
Battery capacity: 9.6Ah/142Wh
Limited charging voltage: 16.8V
Charging temperature range: 0°C~ 45°C
Discharging temperature range: -20°C~ 60°C
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...