14 Probes and Biopsy
172
Basic User Manual
To disinfect the endocavitary probe
You should disinfect the endocavitary probe before and after each use.
Perform the following steps:
1.
Clean the probe.
For details, refer to Section 14.1.3 Cleaning the Probe.
2.
Perform the high-level disinfection to the endocavitary probe as shown in Table 14-2
until the recommended soaking period expires.
Or select the recommended disinfectant (see D.3 Disinfectant) to disinfect the probe
by following the instruction provided by disinfectant’s manufacturer.
3.
Rinse the probe with the running sterile water to remove the residual disinfectant
completely.
4.
Dry the probe with a lint-free soft dry cloth.
Table 14-2 Recommended Method for High-level Disinfection
Disinfectant
Manufacturer Active Ingredients Level for Active
Ingredient
Contact Type Contact Period
Cidex™ Activated
Dialdehyde Solution
J&J
Glutaraldehyde
2.4%
Soak
45-50 minutes
Resert XL HLD
STERIS
H
2
O
2
2.0%
Soak
8 minutes
To sterilize the probe
You should sterilize the intraoperative, biopsy use or blood contacting probe after each use.
Perform the following steps:
1.
Clean the probe.
For details, refer to Section 14.1.3 Cleaning the Probe.
2.
Sterilize the probe as shown in Table 14-3 until the recommended soaking period
expires.
Or select the recommended disinfectant (see D.3 Disinfectant) to sterilize the probe
by following the instruction provided by disinfectant’s manufacturer.
3.
Rinse the probe with the running sterile water to remove the residual sterilant
completely.
4.
Dry the probe with a lint-free soft dry cloth.
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...