14 Probes and Biopsy
Basic User Manual
171
Classification Definition
Level of
Disinfection
Application
Critical
A device that enters normally sterile
tissue or the vascular system.
Sterilization
Intraoperative,
biopsy use or
blood contacting
probe
Semi-critical
A device that comes into contact with
intact mucous membranes and does not
ordinarily penetrate sterile tissue.
High
Endocavitary
probe
Noncritical
Devices that do not ordinarily touch the
patient or touch only intact skin.
Medium or low Body surface
probe
Immersion Requirements
Do not immerse the probe beyond its junction line.
To disinfect the body surface probe
You should disinfect the surface probe after each use.
Perform the following steps:
1.
Clean the probe.
For details, refer to Section 14.1.3 Cleaning the Probe.
2.
Perform the medium-level disinfection to the surface probe as shown in Table 14-1
until the recommended disinfection period expires.
Or select a recommended disinfectant (see D.3 Disinfectant) to disinfect the probe by
following the instruction provided by disinfectant’s manufacturer.
3.
Rinse the probe with the fresh running water to remove the residual disinfectant
completely.
4.
Dry the probe with a lint-free soft dry cloth.
Table 14-1 Recommended Method for Medium-level Disinfection
Disinfectant
Manufacturer Active Ingredients
Level for Active
Ingredient
Contact Type Contact Period
70% Isopropyl
alcohol
All
70% Isopropyl alcohol 70%
Spray/Wipe <10 minutes
T-spray II
Pharm. Inc.
Quat.Ammonia
/
Spray/Wipe <10 minutes
T-spray
Pharm. Inc.
Quat.Ammonia
/
Spray/Wipe <10 minutes
Summary of Contents for P25 EXP
Page 1: ...User Manual P25 EXP Ultrasound System Version 1 4...
Page 5: ......
Page 12: ...Contents VIII Basic User Manual D 3 Disinfectant 200 AppendixEAcousticOutputData 201...
Page 48: ...This page is intentionally left blank...
Page 168: ...This page is intentionally left blank...
Page 176: ...This page is intentionally left blank...
Page 200: ...This page is intentionally left blank...
Page 213: ...201 Appendix E Acoustic Output Data Please consult the accompanying CD...