Ch.6 Web Maintenance Mode
< 176 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Item
Default
Descriptions
ERROR LEVEL
CH1~16 : STANDARD
Sets the method of returning DICOM errors to diagnostic devices.
Do not make changes to this setting unless instructed to do so.
<STANDARD>
: DICOM N-GET responses:
Tray empty
→
WARNING
Maintenance in process
→
FAILURE
Error
→
FAILURE
<FAILURE>
: DICOM N-GET responses
Tray empty
→
FAILURE
Maintenance in process
→
FAILURE
Error
→
FAILURE
<N-ACTION>
: No error returned as DICOM N-GET response.
Error returned upon N-ACTION response.
The relationship between communication between DRYPRO 793 and external devices according to
DICOM error level settings is shown below.
Setting
DICOM Communication
Print Job
DRYPRO 793 Status Displayed
Message
Registration
on External Diagnostic Device
STANDARD
NORMAL
α
READY
FAILURE
NORMAL
α
READY
N-ACTION
NORMAL
α
READY
In warming up, ready to print and tray empty statuses
(When the film empty notification is set to "OFF", including when the film is empty.)
Setting
DICOM Communication
Print Job
DRYPRO 793 Status Displayed
Message
Registration
on External Diagnostic Device
STANDARD
FAILURE
β
CHECK PRINTER
FAILURE
FAILURE
β
CHECK PRINTER
N-ACTION
NORMAL
α
READY
In user/service maintenance mode active statuses
Setting
DICOM Communication
Print Job
DRYPRO 793 Status Displayed
Message
Registration
on External Diagnostic Device
STANDARD
WARNING
β
FILM EMPTY
FAILURE
FAILURE
β
FILM EMPTY
N-ACTION
NORMAL
χ
READY
In tray empty status (With tray empty notification set to "ON.")
Setting
DICOM Communication
Print Job
DRYPRO 793 Status Displayed
Message
Registration
on External Diagnostic Device
STANDARD
FAILURE
β
ERROR
FAILURE
FAILURE
β
ERROR
N-ACTION
NORMAL
χ
READY
Error status
α
: Jobs may be registered.
β
: In cases of diagnostic devices that ignore the DRYPRO 793 status, job registration is possible due to
continuing DICOM communication.
In cases of diagnostic devices that accept the DRYPRO 793 status, job registration is not possible due to
discontinuation of DICOM communication.
χ
: Job registration not possible due to discontinuation of communication by DRYPRO 793 according to DICOM
communication (N-ACTION response).