Ch.3 Setup
< 116 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
1
Switch to the service maintenance mode.
p.86 Switching to
the Service
Maintenance
Mode
2
Touch the [PRINT] button.
- Print setting buttons will be displayed.
3
Touch the [PRINT CONDITION] button.
- An input device selection screen will be displayed.
2
3
Item
Default
Descriptions
DEFAULT DMIN/DMAX
CH01~16 : DICOM
Sets acquisition target of maximum/minimum density values used
only for image processing.
"DICOM"
: The value specified by the diagnostic device is
used.
"P-CON"
: Maximum/minimum density data included in "LUT
SETUP" is used. The LUT used is specified from
the diagnostic device.
DICOM PRIORITY
CH01~16 : OFF
Sets acquisition of parameters used for printing.
"ON"
: Uses values input in print condition settings.
"OFF"
: Uses values input in the diagnostic device.
The following items may be set.
- LUT
- SMOOTHING
- DENSITY MAX
- DENSITY MIN
- POLARITY (NEGA/POSI)
- ORIENTATE
- TRIM
- BORDER (Background density)
- SORTER (Sorter destination bin)
4
Touch the button corresponding to the channel
to be set.
- The print condition setting screen (1/7) will be
displayed.