Ch.1 Pre-installation Information
< 20 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
1.3
Installation Preparations
The following items should be checked before proceeding with installation.
1.3.1
1Checking Installation Conditions
Check to ensure that the place of installation meets the following requirements.
Checking the Place of Installation
- There must be no risk of exposure to water.
- There must no risk of exposure to the harmful effects of high pressure, extreme
temperatures, humidity, poor ventilation, direct sunlight, dust, salt, combustible
gases or sulphur.
- There must be no risk of tilting, exposure to vibration or shock (including transport).
- There must be no chemicals must be stored at the place of installation.
- There must be no risk of gaseous emissions.
- Ventilation must be sufficient to disperse any odours produces by DRYPRO 793.
- There must be sources of noise in the vicinity.
- There must be sufficient space to ensure that there will be no risk of tripping over or
pulling on power or ethernet cables.
- There must be sufficient space to ensure that air inlets and outlets at the side and
rear of the unit are not obstructed.
- The installation space must meet the dimensional requirements shown below.
1000mm minimum
100mm minimum
500mm minimum
DRYPRO
793
Front of
Main Body
500mm minimum
Installation Space
630mm
740mm
195mm
460mm
With upper and lower front covers open
(to replace deodorant filter or resolve jams)
With supply tray open (to load films)
With ejection panel open (to resolve jams)
Normal
When installing exhaust duct and clamp
675mm
54mm
26mm
With left/right panels open (to resolve jams)
62mm
580mm
62mm
460mm
Maximum Space Requirements for DRYPRO 793