
Ch.4 Checking Image Quality
< 144 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
1
Print out one calibration sheet.
1)Invoke the menu mode by touching the menu key.
2)Touch the [CALIBRATION] button on the print
setting menu screen.
- The tray selection screen will be displayed.
3)Select the tray in which the film type to be
calibrated is loaded.
4)After touching the [Print] button, touch [Yes] on the
confirmation screen to print the calibration sheet.
- The calibration sheet print job will be registered
and display returned to the maintenance mode
(print setting) screen.
2
3
4
4.2.4 Procedure for Calibration of Internal Densitometer
The procedure for calibration of the internal densitometer is described below.