Ch.3 Setup
< 104 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Item
Default
Descriptions
LOGIN PASSWORD
ON
Sets the method of moving to the maintenance mode with user
privilege.
For details of differences in DRYPRO 793 operation resulting from
log-in password ON/OFF settings, refer to the operation manual.
"ON"
: Passwords are requested separately for each
user.
"OFF"
: Passwords are not requested.
LANGUAGE SELECT
JAPANESE
Selects the language used on the operation panel and in the Web
maintenance mode.
Selection may be made from the following languages:
"ENGLISH"
"GERMAN"
"FRENCH"
"SPANISH"
"ITALIAN"
"PORTUGUESE"
"SCANDINAVIA"
"JAPANESE"
"CHINESE"
LOG WRITE MODE
NORMAL
Sets the log output status.
"NORMAL"
: Normal mode
"DETAILS"
: Detailed mode
Unless otherwise instructed, this should be set to "NORMAL."
MEMORY TYPE
(Display only)
512
Displays the print memory size mounted in DRYPRO 793. Changes
in the print memory may be checked from the displayed value.
"512"
: 512MB print memory mounted.
"1024"
: 1GB print memory mounted.
TOTAL OPERATING TIME
0
Displays the total DRYPRO 793 operating time.
(Display only)
TOTAL PRINTED SHEETS
0
Displays the total number of sheets printed by DRYPRO 793.
(Display only)
VERSION
Differs depending
Displays current software versions operating in DRYPRO 793.
(Display only)
on date of shipment
Version numbers after software upgrades may be checked from this
display.
MAC ADDRESS
Differs depending
Displays the DRYPRO 793 MAC address.
(Display only)
on the model
Used for reference when WOL (Wake ON LAN) is set in the Regius
console.
p.127 List of Time
Zones
3.7 System Setup
The procedure for making DRYPRO 793 system operational settings is described below.
The following system settings should be made.
In cases where the optional memory is installed, check that "1024" is shown in
the display.(No setting changes are necessary.)
p.177 System Setup
(Web Mainte. Mode)
Changing and saving system settings will result in automatic re-initialization of
DRYPRO 793.
The MAC address can be checked only from the DRYPRO 793 service
maintenance mode.