Ch.3 Setup
< 98 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Item
Default
Descriptions
CHANNEL USE
CH01 : ON
Setting to determine whether or not to use channels of the selected
CH02 : ON
diagnostic device."ON" or "OFF" may be selected.
CH03 : ON
CH04~16 : OFF
SCU NAME
CH01 : Regius170
Sets the name of the selected diagnostic device.
CH02 : Printlink2
A maximum of 12 alphanumerical characters may be specified.
CH03 : Regius190
The device name specified here is used for display.
CH04~16 : (Blank)
IP ADDRESS
CH01 : 192.168.20.90
Sets the IP address of the diagnostic device.
CH02 : 192.168.20.200
"0.0.0.0" should be specified in cases where N-EVENT REPORT is
CH03 : 192.168.20.90
not used.
CH04~16 : 0.0.0.0
HOST NAME
CH01~16 :(Blank)
Sets the host name of the diagnostic device.
This should be left blank in cases where N-EVENT REPORT is not
used.
A maximum of 16 alphanumerical characters may be specified.
PORT NUMBER
CH01 : 6000
Specifies the diagnostic device port number.
CH02 : 6001
"0" should be specified in cases where N-EVENT REPORT is not
CH03 : 6000
used.Settings may be made within a range of 0 – 65535.
CH04~16 : 0
AE TITLE
CH1 : KC_CS1L_U001
Specifies the diagnostic device AE title.
CH2 : KC_PLNK3_U001
A maximum of 16 alphanumerical characters may be specified.
CH3 : KC_CS3L_U001
Alphabetical settings are case sensitive.
CH04~16 : (Blank)
This should be left blank in cases of external devices used in the
expanded port specification.
ERROR LEVEL
CH1~16 : STANDARD
Sets the method of returning DICOM errors to diagnostic devices.
Do not make changes to this setting unless instructed to do so.
<STANDARD>
: DICOM N-GET responses:
Tray empty
→
WARNING
Maintenance in process
→
FAILURE
Error
→
FAILURE
<FAILURE>
: DICOM N-GET responses
Tray empty
→
FAILURE
Maintenance in process
→
FAILURE
Error
→
FAILURE
<N-ACTION>
: No error returned as DICOM N-GET response.
Error returned upon N-ACTION response.
The default settings for each of the DICOM SCU items are based on connection of
CS-1/CS-2 (CH1), Printlink2 (CH2) , CS-3(CH3)(one of each connected).
3.6 SCU Setup
The procedure for making SCU settings (DICOM information for diagnostic devices connected to DRYPRO
793) is described below.
The following SCU settings should be made.
p.173 SCU Setup
(Web Mainte. Mode)
未使用の入力装置の「入力装置 使用」は必ず「OFF」に設定してください。