Ch.4 Checking Image Quality
< 148 >
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
4.3 Test Printing
After completion of calibration of the DRYPRO 793 internal densitometer, a test printing should be carried out
to check image quality.
For installation purposes, two flat patterns and one SMPTE pattern should be printed out to check image
quality.
- Flat pattern (Density: 1.00)
- Flat pattern: (Density: 1.50)
- SMPTE pattern
1
Check to ensure that film is loaded in each tray.
- For details of he procedure for loading film, refer to
the operation manual.
2
Print out two flat patterns from supply tray 1.
1)Touch the menu to invoke the menu mode.
2)log-in to the service maintenance mode using
service privilege.
3)Touch the [TEST PRINT] button under the [PRINT]
item.
- The test print selection screen will be displayed.
4)Touch the [FLAT] button.
- The condition selection screen will be displayed.
4.3.1 Printing Flat Patterns
To check for scratches, line and density inconsistency, two flat patterns should be printed out for
each tray.
This test print should be implemented for each tray.
3
4