Default
Descriptions
FILM SIZE
Tray-1 : 14x17
Sets film sizes for film loaded in each of the trays.
Tray-2 : 14x17
"14x17"
: 14inch x17inch film
Tray-3 : 8x10
"14x14"
: 14inch x14inch film
"11x14"
: 11inch x14inch film
"10x12"
: 10inch x12inch film
"8x10"
: 8inch x10inch film
FILM TYPE
Tray-1 : B
Sets the film type for film loaded in each tray.
Tray-2 : B
"B"
: Blue base film
Tray-3 : M
"C"
: Clear base film
"M"
: Mammography film
COUNT
Tray-1 : 0
Sets the film count for film loaded in each tray.
Tray-2 : 0
Settings may be made within a range of 0 – 126.
Tray-3 : 0
Since the product is shipped with no film loaded, the setting in the
shipping configuration is "0".
In the event that film has been loaded prior to this point, the loaded
film count will be displayed.
TRAY USE
Tray-1 : ON
Sets the usage status of each tray.
Tray-2 : OFF
"ON"
: Tray used.
Tray-3 : OFF
"OFF"
: Tray not used.
SHEETS OF PACKAGE
Tray-1 : 126
Sets the film package count for each tray.
Tray-2 : 126
The value printed on the film box +1 should be set.
Tray-3 : 126
(This is to allow for the calibration sheet used when film is loaded.)
Settings may be made within a range of 0 – 126.
TRAY SERIES USE
ON
Sets operation for use of the same film size/type in multiple trays.
"ON"
: Multiple trays operate as one tray.
1つのトレイのフィルムが無くなっても、別のトレ
イに同一フィルムサイズ/タイプのフィルムが残存
している場合、EMPTY状態にはなりません。
"OFF"
: Multiple trays operate independently.
別トレイに同一フィルムサイズ/タイプのフィルム
が残存していても、EMPTY状態になります。
TRAY SERIES SET
OFF
Sets the film set operation mode valid when all trays are empty.
"ON"
: Film set operation is continued.
"OFF"
: Film set operation is not continued.
Display is returned to the normal screen.
SHEETS IN THE TRAY
1
Sets the number of the trays mounted.
"1"
: 1-tray specification
"2"
: 2-tray specification
"3"
: 3-tray specification
3.4 Film Setup
The procedure for making film size and type settings for film output by DRYPRO 793 is described below.
Although the film sizes and types for DRYPRO722/751/752 are read automatically from the film package, these settings must each
be input in the DRYPRO 793 maintenance mode.
The following settings must be made.
p.170 FILM SETUP
(Web Mainte. Mode)