Calibration Procedures
6
Adjustments procedures
U3606A User’s and Service Guide
199
DC voltage gain adjustment procedure
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Press
to enter the DC voltage gain calibration. The DC annunciator
will be illuminated.
2
The primary display will show the uncalibrated value and the upper
secondary display will show the reference value of the adjustment item.
3
Configure each adjustment item shown in
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
AC current
•
1 mA
•
10 mA
•
100 mA
•
1000 mA
0.9 to 1.1
×
full scale
Capacitance
•
0.4 nF
•
1 nF
•
10 nF
•
100 nF
•
1
μ
F
•
10
μ
F,
•
100
μ
F
•
100
μ
F
•
10000
μ
F
0.9 to 1.1
×
full scale
Table 6-2
Valid gain adjustment input values (continued)
Function
Range
Valid amplitude
input values
Page
V
N O T E
If the zero offset adjustment procedure has been recently performed prior
to the DC voltage gain calibration procedure, the adjustment item “Short”
can be omitted.
N O T E
Always complete tests in the same order as shown in
.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...