![Keysight U3606A User'S And Service Manual Download Page 89](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516089.webp)
Digital Multimeter Operation
2
Math Operations
U3606A User’s and Service Guide
67
Hold
The refresh hold feature allows you to capture and hold a reading, within
the specified variation and threshold values, on the front panel display.
This is useful in situations when you want to take a reading, remove the
test probes, and have the reading remain on the display.
When a stable reading is detected, the instrument beeps once (if the
beeper is enabled), and holds the reading on the primary display. You can
select the variation through the utility menu. The default settings is 10% of
the full scale.
A new reading value is updated in the primary display when the variation
of the measured value exceeds the variation preset in the utility menu.
The instrument beeps once (if the beeper is enabled) when a reading
value is updated. See
on page 130 for more
information on configuring the refresh hold settings.
For voltage, current, and capacitance measurements, the reading value will
not be updated when the reading falls below the threshold preset in the
utility menu.
For continuity and diode tests, the reading value will not be updated
when an open state is detected.
Set the variation to “OFF” in the utility menu to enable data hold. See
on page 131 for more information. In data hold mode,
the reading is not updated even if the input signal value changes. The
reading held will remain on the display until you exit the hold mode.
N O T E
If the reading value is unable to reach a stable state (when exceeding the
preset variation), the reading value will not be updated.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...