5
Verification and Performance Tests
Performance Verification Tests
170
U3606A User’s and Service Guide
6
Operate the electronic load in constant resistance mode, and set its
resistance to 2.5
Ω
. Ensure that the U3606A is operating within the
specified limit and protection values. If not, adjust the electronic load
by reducing the resistance so that the voltage drops slightly until the
U3606A is operating within the specified limit and protection values.
7
Adjust the AC voltage source to the low line voltage limit. Divide the
voltage drop (reading on the digital multimeter) across the current
monitoring resistor by its resistance. Record the output current reading
on the digital multimeter.
8
Within a few seconds after
, adjust the AC voltage source to the
high line voltage limit. Record the output current reading on the digital
multimeter.
9
Compare the difference between the measurement results to the
appropriate test limits shown in
CC noise effect
This test measures the RMS output current in the frequency range from
20 Hz to 1 MHz.
1
Turn off the instrument and connect a 2.5
Ω
fix resistive load between
the front panel
(red) and
(black) output terminals (see
).
2
Connect an AC/DC converter to the fix resistive load and a 50
Ω
feed- thru termination to the AC/DC converter. Connect a differential
amplifier to the 50
Ω
feed- thru termination (see
3
Configure the AC/DC converter as follows:
i
Set the range to 5 A/V.
ii
Set the coupling to AC.
Table 5-14
Constant current source effect verification test
Input AC line
Low line
current limit
High line
current limit
Error from
nominal 1 year
100
90 V
ac
110 V
ac
±0.0015 A
115
104 V
ac
127 V
ac
±0.0015 A
230
207 V
ac
253 V
ac
±0.0015 A
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...