5
Verification and Performance Tests
Performance Verification Tests
162
U3606A User’s and Service Guide
Output verification test
This test verifies that the output functions (constant voltage and constant
current) are within specifications. Note that the measurement values over
the remote interface should be identical to those displayed on the front
panel.
CV programming and readback accuracy
This test verifies that the constant voltage programming and readback
accuracy are within published specifications.
1
Turn off the instrument and connect a digital multimeter between the
front panel
(red) and
(black) output terminals (see
2
Turn on the instrument. Press
Voltage
to select the constant voltage
mode. Ensure that the S1 (30 V/1 A) range is selected (the S1
annunciator is illuminated). Press
Shift
>
Range
again if it is not.
3
Output each value in the order shown below. Record the output voltage
reading on the digital multimeter. Compare the measurement results to
the appropriate test limits shown in
N O T E
You should consider programming the U3606A over the remote interface
for this test to avoid round off errors. See the
U3606A Programmer’s
Reference
for more information on remote interface programming.
Table 5-9
Constant voltage programming and readback accuracy verification test
Output voltage
Range
Error from nominal 1 year
0 V
30 V
±0.005 V
30 V
30 V
±0.02 V
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...