6
Calibration Procedures
Adjustments procedures
198
U3606A User’s and Service Guide
Valid gain adjustment input values
Gain adjustment can be accomplished using the following input values.
Table 6-2
Valid gain adjustment input values
Function
Range
Valid amplitude
input values
Page
DC voltage
•
100 mV
•
1 V
•
–1 V
•
10 V
•
100 V
•
1000 V
0.9 to 1.1
×
full scale
AC voltage
•
10 mV
•
100 mV
•
1 V
•
10 V
•
100 V
•
750 V
0.9 to 1.1
×
full scale
Frequency
Autorange/1 kHz
Input
¥
100 mV
rms
,
900 Hz to 1100 Hz
Resistance
(2-wire)
•
100
Ω
•
1 k
Ω
•
10 k
Ω
•
100 k
Ω
•
1 M
Ω
•
10 M
Ω
0.9 to 1.1
×
full scale
DC current
•
10 mA
•
100 mA
•
1000 mA
0.9 to 1.1
×
full scale