Calibration Procedures
6
Adjustments procedures
U3606A User’s and Service Guide
205
DC current gain adjustment procedure
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Press
to enter the DC current gain calibration. The DC annunciator
will be illuminated.
2
The primary display will show the uncalibrated value and the upper
secondary display will show the reference value of the adjustment item.
3
Configure each adjustment item shown in
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input (see
“Entering adjustment values” on
7
Press
to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
1000.00
Ω
1000
Ω
100.000
Ω
100
Ω
Table 6-6
Resistance gain adjustment
Adjustment item
Input resistance
I
N O T E
If the zero offset adjustment procedure has been recently performed prior
to the DC current gain calibration procedure, the adjustment item “OPEn”
can be omitted.
N O T E
Always complete tests in the same order as shown in
.
O U T
S B Y
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...