Verification and Performance Tests
5
Performance Verification Tests
U3606A User’s and Service Guide
171
4
Connect another 50
Ω
feed- thru termination to the differential amplifier
output and an RMS voltmeter to the 50
Ω
feed- thru termination (see
).
5
Turn on the instrument. Press
Current
to select the constant current
mode. Ensure that the S2 (8 V/3 A) range is selected (the S2
annunciator is illuminated). Press
Shift
>
Range
again if it is not.
6
Set the output current to the full rated value (3 A for S2 range) and
voltage to full scale. Enable the output.
7
Configure the differential amplifier as follows:
i
Set to the AC mode (positive and negative) to remove the DC
component.
ii
Set to the differential mode.
iii
Set the gain to
×
10.
iv
Set the attenuation to 1.
v
Set the low pass filter to 1 MHz bandwidth limit to filter out input
signals containing higher frequencies.
vi
Set to the zero precision current generator.
vii
Set the input impedance to 1 M
Ω
.
8
Configure the RMS voltmeter as follows:
i
Set the high pass filter to 10 Hz.
ii
Enable AC coupling.
9
Obtain the maximum rms current measurement as indicated in the RMS
voltmeter. Divide the value by 10 and then multiply by 5 to get the
constant current rms noise measurement. The result should not exceed
0.001 A
rms
.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...