DC Power Supply Operation
Square-wave Output
92
U3606A User’s and Service Guide
Setting the pulse width
1
To set the square- wave pulse width, press
until the pulse width
value is displayed in the lower secondary display. Ensure that the
square- wave annunciator (
) on the front panel is flashing. (Press
again if it is not.)
2
Press
or
to step through the available duty cycle values.
The pulse width can be stepped through 256 steps, where each step is
1/(256 × frequency) more than the previous step. The display will
automatically adjust to a 5 digit resolution (ranging from 9.9999 ms to
9999.9 ms)
3
Press
again
or
Shift
>
Save
to save the changes.
When all the appropriate parameters have been set, press
to generate
the square wave.
Square-wave pulse width
N O T E
•
Pressing
again will save the changes, and display the next
square-wave parameter. (While still in the edit mode.)
•
Pressing
Shift
>
Save
will save the changes, and exit the edit mode.
•
Changing the square-wave pulse width value will affect the
square-wave frequency and duty cycle values as they are inter-related.
O U T
S B Y
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...