Gain verification test
This test checks the full- scale reading accuracy of the instrument.
Verification checks are performed only for those functions and ranges with
unique gain calibration constants.
DC voltage gain verification test
1
Connect the calibrator to the front panel
V
(red) and
LO
(black) input
).
2
Press
to select the DC voltage function. The DC annunciator will
be illuminated.
3
Select each range in the order shown below. Provide the indicated input
voltage. Compare the measurement results to the appropriate test limits
shown in
. (Be certain to allow for appropriate source settling
when using the Fluke 5520A.)
Table 5-3
DC voltage gain verification test
Input voltage
Range
Error from nominal 1 year
100 mV
100 mV
±0.033 mV
1 V
1 V
±0.0003 V
10 V
10 V
±0.003 V
100 V
100 V
±0.03 V
1000 V
1000 V
±0.3 V
V
C A U T I O N
Set the calibrator output to 0 V before disconnecting from the
instrument input terminals.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...