8
Characteristics and Specifications
Digital Multimeter Specifications
244
U3606A User’s and Service Guide
Digital Multimeter Specifications
Specification assumptions
•
Specifications stated are for 60- minutes warm- up and 5½ digit
resolution.
•
One- year calibration cycle, with calibration temperature of 18 °C to
20 °C.
•
Operating temperature at 18 °C to 28 °C (64.4 °F to 82.4 °F).
•
Accuracy is expressed as: ±(% of r % of range)
•
Temperature coefficient: Add [0.1
×
(the applicable accuracy) / °C] for 0
°C to 18 °C and 28 °C to 55 °C.
•
Relative humidity (RH) up to 80% at 30 °C, proportional to 50% for
30 °C to 55 °C.
DC specifications
Table 8-1
DC accuracy specifications ± (% of r % of range)
Function
Range
[1]
Test current or
burden voltage
24 Hours
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
DC voltage
19.9999 mV
-
0.012 + 0.04
0.015 + 0.04
0.025 + 0.04
0.0015 + 0.0005
100.000 mV
-
0.012 + 0.008
0.015 + 0.008
0.025 + 0.008
0.0015 + 0.0005
1.00000 V
-
0.012 + 0.005
0.015 + 0.005
0.025 + 0.005
0.0010 + 0.0005
10.0000 V
-
0.012 + 0.005
0.015 + 0.005
0.025 + 0.005
0.0020 + 0.0005
100.000 V
-
0.012 + 0.005
0.015 + 0.005
0.025 + 0.005
0.0015 + 0.0005
1000.00 V
-
0.012 + 0.005
0.015 + 0.005
0.025 + 0.005
0.0015 + 0.0005
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...