![Keysight U3606A User'S And Service Manual Download Page 104](http://html.mh-extra.com/html/keysight/u3606a/u3606a_users-and-service-manual_1961516104.webp)
3
DC Power Supply Operation
Protection Functions
82
U3606A User’s and Service Guide
5
Alternatively, press
Shift
>
Exit
or
Shift
>
Protect
to exit the edit mode
without saving.
Checking the OCP operation
To check OCP operation, raise the regulated output voltage slowly. Watch
the current drawn by the load as it reaches near the trip point. Then very
gradually increase the output voltage using the directional keys until the
OCP circuit trips. This will disable the U3606A output, cause the CV
annunciator to flash, and the OC and Error annunciator to illuminate.
The “triP” message appears on the display after a few seconds of
inactivity.
N O T E
•
The OCP feature is always enabled by default whenever you select
constant voltage mode. You can disable the OCP by disabling the output
protection state from the utility menu. See
output protection state,” starting on page 122 for more information.
•
If the OCP value is set to a lesser value than the OC value, the OC value
will be adjusted to equal the OCP value.
•
The OCP value is not applicable for square-wave output.
•
The OCP value is limited by the range selected. Press
Shift
>
Range
to
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).
N O T E
Error number 511, “Current output over protection” will be recorded in the
error queue when the OCP circuit trips. To read and clear the error
message, see
on page 114. For a complete list
of all error messages, see
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...