2
Digital Multimeter Operation
Making Measurements
52
U3606A User’s and Service Guide
Checking diodes
1
Press
again (until the diode annunciator is shown on the
display) to make diode measurements.
2
Connect the red and black test leads to input terminals
(red) and
LO
(black) respectively.
3
Connect the other end of the red test lead to the positive terminal
(anode) of the diode and the black test lead to the negative terminal
(cathode). Refer to
4
Read the display.
N O T E
The resolution for diode tests is fixed to 3½ digits.
Table 2-11
Diode function summary
Item
Description
Measurement method
0.83 mA ± 0.2% constant current source
Audible tone:
•
Continuous beep when level is below +50 mV DC
•
Single tone for normal forward-biased diode or semiconductor
junction where 0.3 V
¤
reading
¤
0.8 V
Input protection
1000 V
rms
on all ranges, < 0.3 A short circuit
Diode annunciator
N O T E
The cathode of a diode is indicated with a band.