Digital Multimeter Operation
2
Making Measurements
U3606A User’s and Service Guide
53
5
Reverse the probes and measure the voltage across the diode again
(refer to
). Assess the diode according to the
following guidelines:
•
A diode is considered good if the multimeter displays “OPEn” in
reverse bias mode.
•
A diode is considered shorted if the multimeter displays
approximately 0 V in both forward and reverse bias modes, and the
U3606A beeps continuously.
•
A diode is considered open if the multimeter displays “OPEn” in both
forward and reverse bias modes.
N O T E
The U3606A can display diode forward bias of up to approximately
1.2 V. The forward bias of a typical diode is within the range of 0.3 V to
0.8 V.