Digital Multimeter Operation
2
Triggering the Multimeter
U3606A User’s and Service Guide
71
Remote interface triggering
Immediate triggering
In the immediate trigger mode, the trigger signal is always present. When
you place the multimeter in the wait- for- trigger state, the trigger is issued
immediately. This is the default trigger source for the U3606A
Multimeter|DC Power Supply.
Remote interface operation:
•
The
TRIGger:SOURce IMMediate
command selects the immediate
trigger source:
•
After selecting the trigger source, you must place the instrument in the
“wait- for- trigger” state using the
INITiate[:IMMediate]
or
READ?
command. A trigger will not be accepted from the selected trigger
source until the instrument is in the “wait- for- trigger” state.
Refer to the
U3606A Programmer’s Reference
for the syntax and complete
descriptions for these commands.
Software (bus) triggering
The bus trigger mode corresponds in function as the
mode, except the trigger is initiated by sending a bus trigger command,
after selecting
BUS
as the trigger source.
N O T E
The immediate trigger mode is available from the remote interface only.
N O T E
The bus trigger mode is available from the remote interface only.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...