5
Verification and Performance Tests
Performance Verification Tests
160
U3606A User’s and Service Guide
Resistance gain verification test
1
Connect the calibrator to the front panel
(red) and
LO
(black) input
).
2
Press
to select the resistance function.
3
Select each range in the order shown below. Provide the indicated input
resistance. Compare the measurement results to the appropriate test
limits shown in
. (Be certain to allow for appropriate source
settling when using the Fluke 5520A.)
Table 5-6
AC current gain verification test
Input current
Input frequency
Range
Error from
nominal 1 year
1 mA
1 kHz
10 mA
±0.015 mA
10 mA
1 kHz
10 mA
±0.06 mA
10 mA
1 kHz
100 mA
±0.15 mA
100 mA
1 kHz
100 mA
±0.6 mA
0.1 A
1 kHz
1 A
±0.0015 A
1 A
1 kHz
1 A
±0.006 A
1 A
1 kHz
3 A
±0.008 A
3 A
1 kHz
3 A
±0.018 A
Table 5-7
Resistance gain verification test
Input resistance
Range
Error from nominal 1 year
100
Ω
100
Ω
±0.058
Ω
[1]
1 k
Ω
1 k
Ω
±0.00055 k
Ω
10 k
Ω
10 k
Ω
±0.0055 k
Ω
100 k
Ω
100 k
Ω
±0.055 k
Ω
1 M
Ω
1 M
Ω
±0.00065 M
Ω
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...