Verification and Performance Tests
5
Performance Verification Tests
U3606A User’s and Service Guide
159
AC current verification test
1
Connect the calibrator to the front panel
I
(red) and
LO
(black) input
).
2
Press
to select the AC current function. The AC annunciator will
be illuminated.
3
Select each range in the order shown below. Provide the indicated input
current and frequency. Compare the measurement results to the
appropriate test limits shown in
. (Be certain to allow for
appropriate source settling when using the Fluke 5520A.)
Table 5-5
AC voltage gain verification test
Input voltage
Input frequency
Range
Error from
nominal 1 year
10 mV
1 Hz
100 mV
±0.12 mV
100 mV
1 Hz
100 mV
±0.3 mV
0.1 V
1 kHz
1 V
±0.0012 V
1 V
1 kHz
1 V
±0.003 V
1 V
1 kHz
10 V
±0.012 V
10 V
1 kHz
10 V
±0.03 V
10 V
1 kHz
100 V
±0.12 V
100 V
1 kHz
100 V
±0.3 V
100 V
1 kHz
750 V
±0.95 V
750 V
1 kHz
750 V
±2.25 V
C A U T I O N
Set the calibrator output to 0 V before disconnecting from the
instrument input terminals.
I
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...