4
System Related Operation
Storing and Recalling Instrument States
132
U3606A User’s and Service Guide
Storing and Recalling Instrument States
You can save and recall complete instrument states including all front
panel settings, all math registers, all utility menu settings, and all bus
specific settings.
There are sixteen user storage registers numbered 1 through 16. An
additional state, state 0, is managed by the instrument and stores the last
power- down state. The instrument automatically saves the complete
instrument configuration to State 0 whenever a power- down event occurs.
For remote operation, refer to the
MEMory:STATe:RECall:AUTO
,
*SAV
,
and
*RCL
commands in the
U3606A Programmer’s Reference
.
Storing a state
To store an instrument state:
1
Press
Shift
>
Store
. The Store annunciator will begin flashing.
2
Press
or
until the state number (01 to 16) you want to store to
is shown on the primary display.
3
Press
Store
or
Shift
>
Save
to store to the selected state. The upper
secondary display briefly shows “donE” when the state is successfully
saved.
4
The instrument will then return to normal operation.
N O T E
If you use state 00 to store your current instrument state, it will be
overwritten with the instrument last power-down state when the power is
cycled.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...