2
Digital Multimeter Operation
Making Measurements
46
U3606A User’s and Service Guide
Performing capacitance measurements
The U3606A Multimeter|DC Power Supply calculates capacitance by
charging a capacitor with a known current for a period of time, and then
measuring the voltage.
Measuring tips:
•
For measuring capacitance values greater than 10000
μ
F, discharge the
capacitor first, then select a suitable range for measurement. This will
speed up the measurement time and also ensure that the correct
capacitance value is obtained.
•
For measuring capacitance values greater than 1 mF, a delayed
response should be expected from the front panel display. For remote
interface operation, increase the SCPI query timeout value. (Typically >
10000 ms.)
•
For measuring small capacitance values, press
Null
with the test leads
open to subtract the residual capacitance of the instrument and leads.
C A U T I O N
Disconnect circuit power and discharge all high-voltage capacitors
before measuring capacitance to avoid damaging the U3606A or the
device under test. To confirm that a capacitor has fully discharged, use
the DC voltage measurement. See
“Measuring DC voltage” on page 26
for more information.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...