DC Power Supply Operation
Protection Functions
80
U3606A User’s and Service Guide
Resetting the OVP condition
Use one of the following methods to reset the OVP circuit after it
activates. If the condition that caused the overvoltage shutdown is still
present, the OVP circuit will turn the output off again.
1
When the OVP circuit trips, the U3606A will immediately prompt you to
change the OVP trip level. Use the directional keys to select a higher
OVP trip level and press
Shift
>
Save
or
Voltage
(
Protect
) to save the
changes.
2
Alternatively, press
Shift
>
Exit
or
Shift
>
Protect
to exit the edit mode
without changing the OVP trip level.
3
You may read and clear the error message by accessing the utility
menu. See
on page 114 for more information.
4
Press
to turn the output on.
5
If the OVP continues to trip, try lowering the regulated output current,
or raising the OVP trip level.
N O T E
Error number 510, “Voltage output over protection” will be recorded in the
error queue when the OVP circuit trips. To read and clear the error
message, see
on page 114. For a complete list
of all error messages, see
, “List of Error Messages,” starting on
O U T
S B Y
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...