6
Calibration Procedures
Adjustments procedures
218
U3606A User’s and Service Guide
2
Press
Shift
>
EXT
to enable remote sensing. When the U3606A is
operating in remote sensing mode, the EXT annunciator on the front
panel is illuminated.
3
Perform
through
of the
“CV output adjustment procedure
— front output terminals” on page 213
for the constant voltage output adjustment items.
4
Repeat
through
“CV output adjustment procedure —
front output terminals” on page 213
for the S1 (30 V/1 A) range. Press
Voltage
to select the S1 range (the S1 annunciator is illuminated).
For adjustment item “LOAd”, connect an additional 30
Ω
, 50 W load
across the rear panel
+
and
–
output terminals. Leave the connections
from the output terminals (+ and –) to the input terminals (
V
and
LO
)
intact.
+ S + +
– – –S
N O T E
Do not remove the short bar between the rear panel sense terminals (+S
and –S) are the rear panel output (+ and –) terminals. See
sensing connections” on page 100
for more information on how to
connect the load leads.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...