Verification and Performance Tests
Additional Verification Tests
174
U3606A User’s and Service Guide
Square-wave frequency output verification test
1
Turn off the instrument and connect a universal frequency counter
between the front panel
(red) and
(black) output terminals (see
).
2
Turn on the instrument. Press
to select the square- wave mode.
3
Output each value in the order shown below. Record the output
frequency reading on the universal frequency counter. Compare the
measurement results to the appropriate test limits shown in
Square-wave duty cycle output verification test
1
Turn off the instrument and connect an oscilloscope between the front
panel
(red) and
2
Turn on the instrument. Press
to select the square- wave mode.
3
Output each value in the order shown below. Record the output duty
cycle reading on the oscilloscope. Compare the measurement results to
the appropriate test limits shown in
Table 5-16
Square-wave amplitude output verification test
Output amplitude
Output frequency
Range
Error from
nominal 1 year
30 V
4800 Hz @ 99.60%
30 V
±0.2 V
8 V
4800 Hz @ 99.60%
8 V
±0.2 V
Table 5-17
Square-wave frequency output verification test
Output frequency
Output amplitude
Range
Error from
nominal 1 year
4800 Hz @ 50%
30 V
30 V
±0.25 Hz
4800 Hz @ 50%
8 V
8 V
±0.25 Hz
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...