Calibration Procedures
6
Calibration Security
U3606A User’s and Service Guide
187
4
Apply power and turn on the instrument.
5
The display will show the message “CALib”. The instrument is now
unsecured. (The security codes resets to the factory default value.)
6
Turn off the instrument and remove the power cord.
7
Remove the temporary short installed in
8
Reassemble the instrument.
9
Unsecure the instrument with the default security code “ATU3606A”.
(Use “3606” if you are unsecuring the instrument from the front panel.)
10
You can now enter a new security code if you wish. See
calibration security code” on page 185
. Be sure you record the new
security code.
C A U T I O N
Never calibrate the instrument while the case is open. Reassemble the
instrument before performing any calibration procedures.