3
DC Power Supply Operation
Sweep Functions
94
U3606A User’s and Service Guide
Generating a ramp signal
1
Select your desired output mode by pressing
Voltage
to enable constant
voltage operation or
Current
to enable constant current operation. (The
respective annunciator — CV or CC — will illuminate accordingly.)
2
Select an appropriate range by pressing
Shift
>
Range
. The maximum
amplitude end position will be limited by the range selected. (The
respective annunciator — S1 or S2 — will illuminate accordingly.)
3
If you want to fine- tune the ramp signal parameters, press
Shift
>
Utility
to access the utility menu. See
on page 123 for more information on configuring the ramp
signal parameters. Press
Shift
>
Save
to save the changes and
Shift
>
Exit
to exit the utility menu.
4
Press
Ramp Scan
to select the ramp function. Ensure that the Ramp
annunciator is shown on the display.
5
Press
to generate the ramp signal.
Ramp annunciator
O U T
S B Y
N O T E
The typical ramp dwelling time per step in the U3606A is measured at
300 ms. A longer delay time should be expected if a high number of steps
is programmed. As an example, a 1000 step ramp signal will yield a
10 minute (1000
×
300 ms
×
2) total dwelling time.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...