DC Power Supply Operation
Protection Functions
86
U3606A User’s and Service Guide
Overcurrent limit (OC)
The overcurrent limit prevents the output current flowing through the load
to change beyond the programmed overcurrent limit. If the load effect
exceeds the programmed overcurrent limit, the CV output will be lowered
to maintain the output power across the load. The combination of the OC
and OCP features create a closed loop circuit protection for sensitive load
behaviors.
Setting the OC level
The OC level can be programmed when the output is enabled (OUT) or
disabled (SBY). Ensure that the U3606A is operating in constant voltage
mode. (The CV annunciator is illuminated.)
1
Press
Shift
>
Limit
to enter the OC edit menu.
2
Press
or
to select which digit position to be changed.
3
Press
or
to increase or decrease the value. The OC value is set
to the maximum protection limits by default.
•
1.0000 A when range S1 is selected, and
•
3.0000 A when range S2 is selected.
4
Press
Shift
>
Save
or
Current
(
Limit
) to save the changes. “donE” will be
briefly displayed in the upper secondary display. The display will then
return to normal.
5
Alternatively, press
Shift
>
Exit
or
Shift
>
Limit
to exit the edit mode
without saving.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...