3
DC Power Supply Operation
Remote Sensing
104
U3606A User’s and Service Guide
Enable remote sensing
To configure the U3606A for remote sensing:
1
Turn off the U3606A.
2
Remove the connections between the U3606A sensing (S+ and S–) and
output (+ and –) terminals. Using a shielded two- wire cable, connect
the U3606A sensing terminals to the load as shown in
Observe the polarity when connecting the sensing leads to the load.
3
Turn on the U3606A.
4
Press
Voltage
to select the constant voltage mode. Use the directional
keys to select the appropriate constant voltage value. See
5
Press
Shift
>
EXT
to enable remote sensing. When the U3606A is
operating in remote sensing mode, the EXT annunciator on the front
panel is illuminated.
6
Press
to regulate the output voltage.
7
Press
Shift
>
EXT
again to disable remote sensing.
C A U T I O N
Do not use the shield as one of the sensing conductors. The other end
of the shielded two-wire cable should be left unconnected.
EXT annunciator
O U T
S B Y
N O T E
If the power supply is operated with remote sensing and either the positive
or negative load wire is not connected, an internal protection circuit will
activate and shut down the power supply. To resume operation, turn the
power supply off, connect the open load wire, and turn on the power
supply.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...