1
Getting Started
Product at a Glance
18
U3606A User’s and Service Guide
•
Press
to select the square-wave output. Use the directional keys to set the
voltage amplitude.
•
Press
again to cycle through the duty cycle, pulse width, and voltage amplitude
settings.
•
See
“Square-wave Output” on page 88
for more information.
When the
annunciator is flashing, press
or
to step through the available
frequencies.
While the
annunciator is flashing, press
or
to set the voltage amplitude, or
to step through the available duty cycle values or pulse width values.
Press
Sweep
to cycle through the ramp and scan sweep functions, or to disable the sweep
mode for the selected output (CV or CC).
[4]
See
information.
Press
Shift
>
Limit
to set the overcurrent limit value for the CV output or the overvoltage
limit value for the CC output. See
“Protection Functions” on page 78
for more information.
Press
Shift
>
Protect
to set the overcurrent protection value for the CV output or the
overvoltage protection value for the CC output
[5]
. See
“Protection Functions” on page 78
for more information.
Press
Shift
>
Range
to toggle between range S1 (30 V/1 A) and range S2 (8 V/3 A)
[6]
. See
“Selecting a Range” on page 97
for more information.
Press
to toggle between source output (
OUT
) and source standby (
SBY
). See
“Enabling the Output” on page 98
for more information.
Press
Shift
>
EXT
to enable remote sensing. See
information.
[1] The voltage path is the default path when you select the frequency measurement function. To switch to the current path for
frequency, pulse width, and duty cycle measurements, first press
, then press
Hz ms %
.
[2] The refresh hold variation and threshold values can be configured through the utility menu. See
Operation,” starting on page 107
for more information on the utility menu.
Table 1-2
U3606A keypad functions (continued)
Key
Description
O U T
S B Y
I
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...