DC Power Supply Operation
3
Protection Functions
U3606A User’s and Service Guide
85
N O T E
•
The OV limit feature is always enabled by default whenever you select
constant current mode. You cannot disable the OV limit feature.
•
If the OV value is set to a greater value than the OVP value, the OVP
value will be adjusted to equal the OV value.
•
If the OV value is set to zero, this will result in the output current being
dropped to zero for limiting.
•
The OV value is not applicable for square-wave output.
•
The OV value is limited by the range selected. Press
Shift
>
Range
to
select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).