Characteristics and Specifications
8
Digital Multimeter Specifications
U3606A User’s and Service Guide
249
Duty cycle and pulse width specifications
100 V
10 V
12 V
750 V
100 V
[1] Maximum input for specified accuracy = 10
×
range or 750 V
rms
or 1000 V
dc
Table 8-4
Frequency sensitivity for voltage measurement (continued)
Input range
[1]
Minimum sensitivity (rms sine wave)
20 Hz to 100 kHz
100 kHz to 300 kHz
300 kHz to 1 MHz
Table 8-5
Frequency sensitivity for current measurement
Input range
Minimum sensitivity (rms sine wave)
20 Hz to 10 kHz
10 mA
1 mA
100 mA
10 mA
1.000 A
100 mA
3 A
300 mA
Table 8-6
Duty cycle and pulse width resolution and accuracy
Function
Range
Resolution
Accuracy of full scale
Duty cycle
100.000%
[1]
0.001%
0.3% + 0.2% per kHz
Pulse width
199.999 ms
[2]
0.001 ms
Duty cycle/Frequency
1999.99 ms
0.01 ms
Duty cycle/Frequency
[1] The range is calculated from {10
μ
s
×
frequency
×
100%} to {[1 – (10
μ
s
×
frequency
)]
×
100%}. For example, a 1 kHz signal
can be measured from 1% to 99%.
[2] The positive or negative pulse width must be greater than 10
μ
s. The range of the pulse width is determined by the frequency
of the signal.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...