6
Calibration Procedures
Recommended Test Equipment
180
U3606A User’s and Service Guide
Recommended Test Equipment
The test equipment recommended for the adjustment procedures is listed
below. If the exact instrument is not available, substitute with calibration
standards of equivalent accuracy.
Table 6-1
Recommended test equipment for adjustment procedures
Application
Recommended equipment
Recommended accuracy requirements
Zero calibration
Shorting plug — Dual banana plug with copper wire
short between the two input terminals
-
DC voltage
Calibrator
Fluke 5520A
< 20% of instrument 1 year specification
DC current
Calibrator
Fluke 5520A
< 20% of instrument 1 year specification
Calibrator
Fluke 5520A
< 20% of instrument 1 year specification
AC current
Calibrator
Fluke 5520A
< 20% of instrument 1 year specification
Resistance
Calibrator
Fluke 5520A
< 20% of instrument 1 year specification
Frequency
Function/Arbitrary
waveform generator
Keysight 33250A
< 20% of instrument 1 year specification
Capacitance
Calibrator
Fluke 5520A
< 20% of instrument 1 year specification
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...