Calibration Procedures
6
Adjustments procedures
U3606A User’s and Service Guide
213
CV output adjustment procedure — front output terminals
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Turn off the instrument and connect the front panel
(red) and
(black) output terminals to the
V
(red) and
LO
(black) input
terminals.
2
Turn on the instrument. Press
Voltage
to enter the constant voltage
output calibration. Ensure that the S2 (8 V/3 A) range is selected (the
S1 annunciator is illuminated). Press
Voltage
again if it is not.
3
The primary display will show the uncalibrated value and the upper
secondary display will show the reference level of the adjustment item.
4
Configure each adjustment item shown in
5
Use
or
to select the adjustment item.
S1 (30 V/1 A)
OUt-L
0 A
OUt-H
1 A
LOAd
[1]
1 A with additional
30
Ω
, 50 W load
for information on how to connect the LOAd adjustment item.
Table 6-11
Constant current output adjustment (continued)
Range
Adjustment item
Output current
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...