6
Calibration Procedures
Adjustments procedures
206
U3606A User’s and Service Guide
•
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the DC current gain adjustments using the
verification test” on page 158
.
AC current gain adjustment procedure
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Press
twice to enter the AC current gain calibration. The AC
annunciator will be illuminated.
2
The primary display will show the uncalibrated value and the upper
secondary display will show the reference value of the adjustment item.
3
Configure each adjustment item shown in
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
Table 6-7
DC current gain adjustment
Adjustment item
Input current
OPEn
Input terminals open (remove any test leads or
shorting plugs from the input terminals)
10.0000 mA
10 mA
100.000 mA
100 mA
1000.00 mA
1000 mA
I
N O T E
Always complete tests in the same order as shown in
.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...