6
Calibration Procedures
Adjustments procedures
202
U3606A User’s and Service Guide
Frequency gain adjustment procedure
Review the
“Test Considerations” on page 145
sections before beginning this procedure.
1
Press
Hz ms %
to enter the frequency gain calibration.
2
The primary display will show the uncalibrated value and the upper
secondary display will show the reference value of the adjustment item.
3
Configure each adjustment item shown in
4
Use
or
to select the adjustment item.
5
Apply the input signal shown in the “Input” column of
6
Enter the actual applied input (see
“Entering adjustment values” on
Table 6-4
AC voltage gain adjustment
Adjustment item
Input voltage (V
rms
)
Input frequency
10.000 mV, 1 kHz
10 mV
1 kHz
100.000 mV, 1 kHz
100 mV
1 kHz
1.00000 V, 1 kHz
1 V
1 kHz
10.0000 V, 1 kHz
10 V
1 kHz
100.000 V, 1 kHz
100 V
1 kHz
750.00 V, 1 kHz
750 V
1 kHz
C A U T I O N
Set the calibrator output to 0 V before disconnecting from the
instrument input terminals.
N O T E
Always complete tests in the same order as shown in
.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...