Calibration Procedures
Adjustments procedures
200
U3606A User’s and Service Guide
6
Enter the actual applied input (see
“Entering adjustment values” on
7
Press
to start the adjustment. The display “CALib” in the lower
secondary display starts flashing to indicate that the calibration is in
progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
•
An adjustment failure is indicated by a long beep, the primary
display showing “FAiL” and a calibration error number appearing in
the upper secondary display. Correct the problem and repeat this
procedure.
8
Repeat
through
for each gain adjustment point shown in
.
9
Verify the DC voltage gain adjustments using the
verification test” on page 157
.
Table 6-3
DC voltage gain adjustment
Adjustment item
Input voltage
Short
Dual banana plug with copper wire short
between the two input terminals
100.000 mV
100 mV
1.00000 V
+1 V
–1.00000 V
–1 V
10.0000 V
10 V
100.000 V
100 V
1000.00 V
1000 V
O U T
S B Y
C A U T I O N
Set the calibrator output to 0 V before disconnecting from the
instrument input terminals.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply User s and Service Guide ...
Page 2: ......
Page 12: ...12 U3606A User s and Service Guide ...
Page 128: ...3 DC Power Supply Operation Remote Sensing 106 U3606A User s and Service Guide ...
Page 160: ...4 System Related Operation Remote Operation 138 U3606A User s and Service Guide ...
Page 242: ...6 Calibration Procedures Adjustments procedures 220 U3606A User s and Service Guide ...
Page 291: ...List of Error Messages 9 Error Messages U3606A User s and Service Guide 269 ...