Remote Interface Selection
The multimeter is shipped with both an
HP-IB
(
IEEE-488
) interface
and an
RS-232
interface. Only one interface can be enabled at a time.
The
HP-IB
interface is selected when the multimeter is shipped from
the factory.
The remote interface can be set only from the front-panel.
•
The interface selection is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
•
If you select the
HP-IB
interface, you must select a unique address for
the multimeter. The
HP-IB
address is displayed when you turn on the
multimeter.
•
If you select the
RS-232
interface, you must set the baud rate and
parity for the multimeter. “
RS-232
” is displayed when you turn on the
multimeter.
•
If you select the
RS-232
interface and then set the
HP-IB
address to
the talk only address (31), the multimeter will
send
readings over the
RS-232
interface when in the local mode.
•
There are certain restrictions to be aware of when you are selecting
the remote interface (see also “Programming Language Selection,” on
page 94). The only programming language supported on
RS-232
is
SCPI
.
•
Front-Panel Operation
:
2: INTERFACE (I/O MENU)
See also “To Select the Remote Interface,” on page 162.
SCPI Language
HP 3478A Language
Fluke 8840A Language
HP-IB / 488
X
X
X
RS-232
X
Not allowed
Not allowed
Chapter 3 Features and Functions
Remote Interface Configuration
92
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...