Integration Time
(continued)
•
Front-Panel Operation
: Integration time is set indirectly when you
select the number of digits.
See the table for resolution on page 54.
•
Remote Interface Operation
:
<
function
>:NPLCycles {0.02|0.2|1|10|100|MINimum|MAXimum}
For frequency and period measurements,
aperture time
(or gate time)
is analogous to integration time. Specify 10 ms (4
1
⁄
2
digits),
100 ms
(default; 5
1
⁄
2
digits), or 1 second (6
1
⁄
2
digits).
FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
PERiod:APERture {0.01|0.1|1|MINimum|MAXimum}
Front / Rear Input Terminal Switching
Any measurement made using the front terminals can also be made
using the input terminals on the rear panel. See “The Front Panel at
a Glance,” on page 2, for the location of the front / rear switch.
The input terminals can only be configured from the front panel.
You cannot select the terminals from the remote interface, but you
can query the present setting.
•
The
Rear
annunciator turns on when you select the rear terminals.
•
Remote Interface Operation
: You can query the multimeter to
determine whether the front or rear input terminals are selected.
ROUTe:TERMinals?
returns “
FRON
” or “
REAR
”
Chapter 3 Features and Functions
Measurement Configuration
58
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...