Continuity Threshold Resistance
When measuring continuity, the multimeter emits a continuous tone if
the measured resistance is less than the
threshold resistance
. You can
set the threshold to any value between 1
Ω
and 1000
Ω
.
The threshold resistance is adjustable only from the front panel.
•
The threshold resistance is stored in
non-volatile
memory, and
does not
change when power has been off or after a remote interface reset.
•
The factory setting for the threshold resistance is 10
Ω
.
•
After enabling the continuity function, you can select a different
threshold resistance by pressing
Shift >
(Menu Recall).
2: CONTINUITY (MEAS MENU)
See also “To Test Continuity,” on page 19.
∧
0010 OHM
Chapter 3 Features and Functions
Measurement Configuration
52