Operator Maintenance
This section describes how to replace the power-line and current fuses.
If you need additional information about replacing parts or repairing
the multimeter, see the
Service Guide
.
To Replace the Power-Line Fuse
The power-line fuse is located within the multimeter’s fuse-holder
assembly on the rear panel (
see also page 15
). See the rear panel of the
multimeter for the proper fuse rating. To replace a 250 mAT fuse, order
HP part number 2110-0817. To replace a 125 mAT fuse, order HP part
number 2110-0894.
To Replace the Current Input Fuses
The front and rear current input terminals are protected by two series
fuses. The first fuse is a 3A, 250 Vac, fast-blow fuse and is located on the
rear panel. To replace this fuse, order HP part number 2110-0780.
A second fuse is located inside the multimeter to provide an additional
level of current protection. This fuse is a 7A, 250 Vac, high-interrupt
rated fuse (HP part number 2110-0614). To replace this fuse, you must
remove the multimeter’s case by loosening three screws.
See the Service
Guide for more information on disassembling the multimeter.
With a small flatblade screwdriver, push in on the fuse cap
and rotate it counterclockwise. Remove the fuse cap and fuse.
Chapter 3 Features and Functions
Operator Maintenance
100
Summary of Contents for 34401A
Page 12: ...1 Quick Start 1 ...
Page 26: ...2 Front Panel Menu Operation 2 ...
Page 50: ...3 Features and Functions 3 ...
Page 103: ...4 Remote Interface Reference 4 ...
Page 135: ...SCPI Status System 4 Chapter 4 Remote Interface Reference The SCPI Status Model 135 ...
Page 170: ...5 Error Messages 5 ...
Page 182: ...6 Application Programs 6 ...
Page 196: ...7 Measurement Tutorial 7 ...
Page 214: ...8 Specifications 8 ...