Resolution
(continued)
•
Front-Panel Operation
: Select either the slow or fast mode for each
resolution setting.
The default mode is 5 digits slow.
5: RESOLUTION (MEAS MENU)
See also “To Set the Resolution,” on page 21.
•
Remote Interface Operation
: You can set the resolution using the
following commands.
Specify the resolution in the same units as the measurement
function,
not in number of digits
. For example, for dc volts, specify the
resolution in volts. For frequency, specify the resolution in hertz.
CONF:VOLT:DC 10,0.001
4
1
⁄
2
digits on the 10 Vdc range
MEAS:CURR:AC? 1,1E-6
6
1
⁄
2
digits on the 1 A range
CONF:FREQ 1 KHZ,0.1 Hz
1000 Hz input, 0.1 Hz resolution
VOLT:AC:RES 0.05
50 mV resolution on the ac function
CONFigure:<
function
> {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
MEASure:<
function
>? {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
<
function
>:RESolution {<
resolution
>|MIN|MAX}